ABSTRACT

<p><b>OBJECTIVE</b>To evaluation the performance of a heart-type fatty acid-binding protein (H-FABP) ELISA detection kit in the clinical diagnosis of acute coronary syndrome (ACS).</p><p><b>METHODS</b>Plasma or serum samples from 160 suspected ACS patients hospitalized in General Hospital of PLA were examined using Lanzhou H-FABP reagent kit and Holland H-FABP kit. Correlation of the two kits was evaluated and Kappa test was used to examine the consistency of the results of the two products.</p><p><b>RESULTS</b>The sensitivity of H-FABP diagnosis of ACS detection Lanzhou kit was 91.8%, the specificity was 88.7%, and the total diagnostic rate was 90.42%. The sensitivity of H-FABP diagnosis of ACS detection Holland kit was 90.3%, the specificity was 86.8%, and the total diagnostic rate was 88.75%. The test results showed that two products yielded comparable results (P=0.668, >0.05) with a good consistency (Kappa=0.726, P<0.01).</p><p><b>CONCLUSIONS</b>H-FABP ELISA detection kit produccted by LanZhou biological research institute has a good correlation with H-FABP detection kit produced by HBT company of Holland and has the potential for clinical application.</p>


Subject(s)
Humans , Acute Coronary Syndrome , Diagnosis , Biomarkers , Blood , Enzyme-Linked Immunosorbent Assay , Fatty Acid Binding Protein 3 , Fatty Acid-Binding Proteins , Blood , Reagent Kits, Diagnostic , Reproducibility of Results , Sensitivity and Specificity